Review of Scientific Instruments · 1999 · 44 citations · 10 references
Natural SciencesPhysicsPlasma PhysicsRadiation DetectionSpectroscopyRadiation SpectroscopyX-ray DetectorsX-ray SpectroscopyLinear PolarizationLinear Polarization MeasurementPolarizationSoft X-ray SpectroscopySpectrometer DesignX-ray Emission LinesX-ray Emission SpectroscopyX-ray Line Emission
Line emission from highly charged ions in the soft x-ray region represents an important diagnostic for high-temperature plasmas. The wavelength region of 10–30 Å is occupied by lines from the K-shell transitions in N through Ne and L-shell transitions in Ti through Ni. Crystal spectrometers have previously been used to investigate Fe line radiation under well controlled conditions using the LLNL electron beam ion trap facility in this wavelength region. To complement this work we have instituted a grazing-incidence spectrometer with a variable line-spaced, concave grating and a flat field of focus which is suited for a multichannel detector. Simultaneous observations of the Ne-like Fe16+ n=3→2 transitions were made with this instrument and a flat-crystal spectrometer in order to establish the linear polarization of the Fe XVII lines. We present a description of the grating spectrometer and discuss its performance in comparison with Rowland circle based grazing incidence instruments.
10
Mechanically ruled aberration-corrected concave gratings
Tatsuo Harada, Toshiaki Kita · Applied Optics · 1980 · 245 citations
Laboratory Measurements and Modeling of the FexviiX‐Ray Spectrum
G. V. Brown, P. Beiersdörfer, D. A. Liedahl et al. · The Astrophysical Journal · 1998 · 174 citations