Magnetic Resonance in Medicine · 1991 · 160 citations · 11 references
Natural SciencesRadiologyDiagnostic ImagingDiagnostic RadiologySpectroscopyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Pulse SequencesMagnetic Resonance Signal FormationMagnetic Resonance SpectroscopyMedical ImagingAdiabatic PulsesExcitationPulse Sequence DesignPulse SequencesSpatial InhomogeneitySpin Echo FormationSpin Echo SequencesVariable-rate Selective Excitation
We introduce a selective adiabatic pulse sequence suitable for generating selective spin-echoes for both MR imaging and spectroscopy. The technique is simple; one uses the echo generated by any pair of identical selective adiabatic inversion pulses. The nonlinear phase across the slice is compensated perfectly by the second pi pulse. This compensation is immune to RF inhomogeneity and nonlinearity. For imaging applications, we concentrate on a reduced-power version of the pulse sequence in which time is traded off variably for RF amplitude in the presence of a time-varying gradient. This technique, known as variable-rate excitation, mildly degrades the off-resonant slice profile when applied to amplitude-modulated pulses. We present theoretical explanations and experimental results that show that the variable-rate adiabatic pulses are immune to off-resonant degradation of the magnitude normally encountered in MR imaging.
11
Variable-rate selective excitation
Steven Conolly, Dwight G. Nishimura, Albert Macovski et al. · Journal of Magnetic Resonance (1969) · 1988 · 319 citations
Sharmila Majumdar, Stelios C. Orphanoudakis, Arthur F. Gmitro et al. · Magnetic Resonance in Medicine · 1986 · 206 citations