The effective quality factor at low temperatures in dynamic force microscopes with Fabry–Pérot interferometer detection
36
Citations
30
References
2009
Year
EngineeringMicroscopyMechanical EngineeringOptomechanical SystemInterferometryOptomechanicsMicro-electromechanical SystemMicroscopy MethodCalibrationOptical PropertiesMechanicsEffective Quality FactorInstrumentationNanomechanicsDynamic Force MicroscopesBiophysicsLaser PowerPhysicsOptical SensorsLow TemperaturesScanning Probe MicroscopyApplied PhysicsFabry–pérot InterferometerScanning Force MicroscopyMedicine
The oscillation of a cantilever in ultrahigh vacuum dynamic force microscopy is sometimes measured with the help of a Fabry–Pérot interferometer. We show that the photoinduced forces present in such an interferometer can artificially increase or decrease the effective quality factor of the cantilever. We examine this effect on the basis of a PtIr-coated rectangular silicon cantilever at temperatures between 7.7 and 30 K. By measuring resonance curves we demonstrate that the magnitude of the effective quality factor changes with cantilever-fiber distance, laser power, and temperature.