Applied Spectroscopy · 1997 · 14 citations · 6 references
Natural SciencesChemistryNuclear EngineeringAnalytical ChemistryElemental AnalysisSpectroscopyAtomic Emission SpectroscopyEmission SpectroscopyNuclear SpectroscopyRadiochemical AnalysisSpectral DeconvolutionAtomic Emission SpectrometryBurnup MeasurementCalibration Curve FittingCalibration CurvesFission Product AnalysisInductively Coupled Plasma Atomic Emission SpectroscopyInductively Coupled Plasma SpectrometrySpectrum Deconvolution
The development of a method for the estimation of burnup of nuclear fuels by the determination of 236 U by high-resolution inductively coupled plasma spectrometry has been demonstrated. Linear calibration curves are obtained and a limit of quantification of ∼ 3 atom % burnup established. The precision, in terms of the relative standard deviation (RSD) of the determination, was 3.7% RSD for 1.58 atom % of 236 U. These results were obtained by using a spectrometer with a resolving power of 2.3 × 10 5 and with the application of a spectral deconvolution using a nonlinear curve-fitting algorithm. A complementary method, the determination of 139 La, was also investigated and proved to be accurate (recovery = 101.5%) and precise (RSD = 2.2%) with a limit of quantification of ∼ 0.5 atom % burnup. These methods were developed with the use of model solutions which approximate digests of irradiated fuels but which contained no active fission product species.
6
J. Ignacio García Alonso, Fabrizio Sena, Philippe Arbore et al. · Journal of Analytical Atomic Spectrometry · 1995 · 120 citations