Sensitive Signal ProcessingEngineeringAnalytical MicrosystemsBiomedical EngineeringChemistryBiosensorsReflectivity SpectrumHorseradish PeroxidaseBiosensing SystemsAnalytical ChemistryBioimagingNanosensorMicrofluidicsChemical SensorMolecular ImagingBiophysicsPorous SensorPorous Silicon LayerBiophotonicsOptical SensorsBiomedical DiagnosticsMicrofabricationSensor DesignOptical Sensor
Abstract Orthogonal subspace signal processing algorithms (OSPA) have been developed to extract the optical thickness of a porous silicon layer to within one part in 10 5 from its reflectivity spectrum. This is equivalent to a limit of detection (LOD) of ∼40 pm change in optical thickness for a 3 µm thick layer, or an LOD of 1/2000 of a monolayer coverage with antibodies, of molecular weight 160 k Daltons, within a layer with pores of 100 nm diameter. A large molecule {horseradish peroxidase (HRP), MWt 40 kDa} has been detected at a concentration of 1 µg/ml by measuring its direct binding to anti‐HRP antibodies immobilised within a porous silicon layer. A competitive assay has been demonstrated for the detection of a small molecule {2, 4, 6 trinitrotoluene (TNT), MWt 227 Da} at 10 µg/ml. The projected LODs for HRP and TNT by these assays are 50 ng/ml and 1 µg/ml respectively. (© 2005 WILEY‐VCH Verlag GmbH & Co. KGaA, Weinheim)