Analytical Letters · 2011 · 12 citations · 12 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistryInstrumentationWater Quality EngineeringAnalytical CalibrationAnalytical MeasurementsCalibrationEquipment CalibrationCalibration Curve AnalysisCalibration Method ValidationContinuous FlowContinuous Flow OperationFlow CalibrationInstrumental Calibration ErrorsReagent ConsumptionSystematic Error Calibration
The integrated calibration method (ICM) is a novel calibration approach merging interpolative and extrapolative modes in a single procedure. Thanks to this strategy, chemical analysis can be performed along with a diagnosis of systematic calibration errors that leads to better accuracy of analytical results. The paper describes how the ICM can be adapted to different flow techniques: continuous flow, flow-injection, and sequential-injection ones. Instrumental systems dedicated to each flow technique have been presented and their operation has been explained. The systems were used for UV/VIS chromium determination in water samples and compared with each other in terms of precision and accuracy of the obtained results along with time and reagent consumption.
12
J. Ru̇žička, John Stewart · Analytica Chimica Acta · 1975 · 1.5K citations
Michael Sperling, Zhao‐Lun Fang, Bernhard Welz · Analytical Chemistry · 1991 · 58 citations
Edvan C. Silva, Mário César Ugulino de Araújo, Ricardo S. Honorato et al. · Analytica Chimica Acta · 1996 · 29 citations