Application of a New pH‐Sensitive Electrode as a Detector in Flow Injection Potentiometry

Heike Kahlert, Jens R. Pörksen, İbrahim Işıldak, Müberra Andaç, Murat Yolcu, Jürgen Behnert, Fritz Scholz

Electroanalysis · 2005 · 17 citations · 9 references

Concepts

Abstract

Abstract A new pH sensitive detector for flow‐through potentiometry was developed on the basis of a graphite/quinhydrone composite electrode. The detector was constructed of two polymethylmethacrylate plates between which a 0.1 mm thick sensitive layer is situated. After drilling a hole through the plates, a ring of the sensitive layer is exposed to the solution stream. A conventional calomel reference electrode was placed downstream following the flow‐through sensor. The response behavior in different electrolyte solutions was investigated. The detector shows a Nernstian behavior for injections of hydrochloric acid into a KCl background solution as well as for the steady state response in concentration step experiments using hydrochloric acid and buffer solutions. The response time is sufficiently short for FIA applications ( T 95 between 3 and 5.4 s). The influences of flow rate and injection volume on the detector signal are discussed.

References

9