Review of Scientific Instruments · 1971 · 11 citations · 2 references
Natural SciencesChemistryAnalytical ChemistryMass Spectrometric AnalysisSpectroscopyAccurate Mass MeasurementInstrumental AnalysisMass AnalyzersMass SpectrometryWater ChemistryBoilingDeuterium AbundanceFilm BoilingHydrogen Content AnalysisHydrogen DeterminationHydrogen Isotope AnalysisHydrogen Isotope CompositionSampling
The Leidenfrost effect, characterized by film boiling of a liquid adjacent to a hot metal surface, has been adapted in two forms for sampling water vapor for mass spectrometer analysis to determine the deuterium-to-hydrogen (D/H) ratio. Both a hot probe and a hot plate configuration convert liquid water to vapor without interference from dirt or dissolved material in the liquid phase. Automatic and semiautomatic analysis of the D/H ratio in water has been achieved with a precision of ±0.1 ppm at natural deuterium concentrations. A 2 min cycle time has been used for multiple sampling of bulk water and flowing stream samples.
2