Review of Scientific Instruments · 1995 · 32 citations · 11 references
EngineeringMaterials SciencePhysicsRadiation DetectionRadiation MeasurementRadiation PhysicsDetector TechnologyParticle DetectorsRadiation DetectorsSemiconductor DetectorsX-ray DetectionX-ray DetectorsGermanium DetectorsHigh-purity Germanium DetectorsImpuritiesPulse Height AnalysisSolid-state X-ray Detectors
In the energy range from 1 keV down to a few hundred eV, a newly developed ‘‘ultralow-energy measurable’’ pure-Ge detector for a pulse-height analysis and a current-mode observation has been characterized using synchrotron radiation monochromatized by a grasshopper monochromator at the Photon Factory of the National Laboratory for High Energy Physics (KEK). X-ray measurements in this low-energy region were previously tried out using several types of ‘‘windowless’’ Si(Li) detectors. These detectors, however, had trouble, including temporal variations in the quantum efficiencies because of their detector-surface deteriorations due to various impurities in vacuum chambers. Our pure-Ge detector has a 4000-Å thick polymer window metalized by a 1400-Å thick Al supported by a 100-μm apart Si-ribbed structure. However, for this liquid-nitrogen-cooled detector with the special window, its actual energy-response data are not available at this time. Therefore, the investigations of its characteristics are reported particularly for the purpose of the reliable x-ray observations in such a low-energy regime.
11