Wavelength Modulation for Background Correction in Graphite Furnace Atomic Emission Spectrometry

49

Citations

15

References

1976

Year

Abstract

The use of wavelength modulation for background correction in graphite furnace atomic emission spectrometry is shown to improve detection limits in some cases by several orders of magnitude and to increase its applicability to sample analysis. An HGA-2100 graphite furnace atomizer is used in conjunction with a 3/4-m monochromator equipped for wavelength modulation. A discussion of analytical growth curves, detection limits, scatter, and optimization of signal/noise ratio is presented. Copper is analyzed in botanical samples and the results are compared to analysis by atomic absorption and molecular absorption spectrophotometry.