Applied Optics · 1997 · 36 citations · 13 references
Natural SciencesChemistryMetrologyOpticsSpectroscopyFourier Transform SpectroscopyInfrared InstrumentationInfrared SpectroscopyLaser SpectroscopyOptical SpectroscopyRefractive IndexFourier TransformInfrared OpticsNeutral Density FiltersOptical TransmittanceSilicon WafersSystematic Error AnalysisSystematic Error CorrectionSystematic Error Mitigation
We have investigated the ordinate scale accuracy of ambient temperature transmittance measurements made with a Fourier transform infrared (FT-IR) spectrophotometer over the wavelength range of 2-10 mum. Two approaches are used: (1) measurements of Si wafers whose index of refraction are well known from 2 to 5 mum, in which case the FT-IR result is compared with calculated values; (2) comparison of FT-IR and laser transmittance measurements at 3.39 and 10.6 mum on nominally neutral-density filters that are free of etaloning effects. Various schemes are employed to estimate and reduce systematic error sources in both the FT-IR and laser measurements, and quantitative uncertainty analyses are performed.
13
Infrared refractive index of silicon
David F. Edwards, Ellen Ochoa · Applied Optics · 1980 · 190 citations
Natural Sciences, Materials Engineering, Materials Science +11
Nonlinear Detector Response in FT-IR
D. Bruce Chase · Applied Spectroscopy · 1984 · 100 citations
Natural Sciences, Chemistry, Physics +16
Source of a problem with Fourier transform spectroscopy
D. B. Tanner, R. P. McCall · Applied Optics · 1984 · 26 citations