Performance characteristics of a small side-window photomultiplier in laser single-photon fluorescence decay measurements

William R. Ware, Madhuri Pratinidhi, R. K. Bauer

Review of Scientific Instruments · 1983 · 69 citations · 27 references

Concepts

Abstract

The performance characteristics of the Hamamatsu R928 side-window photomultiplier tube have been evaluated in the picosecond time domain. This phototube is proposed as an attractive solution to the problem of high time resolution when a sync-pumped dye laser is combined with the single-photon time correlation method for determining fluorescence lifetimes and measuring time-resolved emission spectra and polarization anisotropy. It combines excellent timing characteristics in the single-photon mode and a small-to-negligible wavelength effect with high gain, low noise, and low cost. Twice the resolution previously reported has been obtained.

References