Journal of the Optical Society of America · 1953 · 21 citations · 3 references
Natural SciencesChemistryInstrumentationOpticsSpectroscopyOptical InstrumentationOptical InstrumentsOptical SpectroscopyRaman SpectroscopyGrating SpectroscopyPhotomultiplier DetectionPhotomultiplier DetectorsPhotomultiplier TubesRaman InstrumentationRaman SpectraScanning MechanismSpectrometer Design
The emphasis in this paper is on the circuitry aspect of the problem of converting a photographic Raman instrument into a photoelectric Raman spectrometer. The reconstructed instrument was tested by using a scanning-slit mechanism in the focal plane of the camera lens of a grating spectrograph. A photomultiplier tube served as the detector and was operated in a vacuum housing at the temperature of liquid nitrogen. A tetrode electrometer tube in a low gain circuit yielded a nearly linear output which was graphically recorded. A spiral mercury arc provided sufficient illumination to enable reproduction of intensity measurements to within ±1 percent at a scanning rate of about 3 cm−1 per second. An automatic range changing device extends the range of intensities which can be recorded on the normal ten-inch chart.
3