European Surgical Research · 2004 · 12 citations · 26 references
MedicineOncologyRadiobiologySurgical InstrumentationTreatment PlanningAblation TherapyHepatic PerfusionLiver Blood FlowAblation Of Liver TumorsCoagulationLiver Tumor AblationPerfusion Rate OptimizationThermal Ablation
In an attempt to increase necrotic zones in liver tissue by radiofrequency ablation fresh bovine liver was coagulated by means of a needle electrode continuously perfused with NaCl solution. Power output (60 W) and application time (15 min) were kept constant while the perfusion was varied in terms of saline concentration (0.9, 5.85 and 10%) and perfusion rate (40 or 80 ml/h). Our results showed that the use of higher osmolar saline solutions in radiofrequency ablation with perfused needle electrodes did not lead to significantly larger coagulation volumes. By contrast, increasing the perfusion rate produced significantly larger necrotic zones. Doubling the perfusion rate made it possible to reach higher temperatures (>60 degrees C) within significantly shorter time.
26
Small Hepatocellular Carcinoma: Treatment with Radio-frequency Ablation versus Ethanol Injection
Tito Livraghi, S. Nahum Goldberg, Sergio Lazzaroni et al. · Radiology · 1999 · 1.3K citations
Radiofrequency Ablation of Unresectable Primary and Metastatic Hepatic Malignancies
Steven A. Curley, Francesco Izzo, Paolo Delrio et al. · Annals of Surgery · 1999 · 1.1K citations · Full text
Factors influencing the natural history of colorectal liver metastases
R. Stangl, A. Altendorf-Hofmann, RM Charnley et al. · The Lancet · 1994 · 721 citations