Publication | Closed Access
Monolithic 180 nm CMOS Dosimeter for In Vivo Medical Applications
22
Citations
5
References
2013
Year
Medical ElectronicsBiomedical EngineeringWireless Implantable DeviceTreatment VerificationMedical InstrumentationRadiation TestingRadiation MedicineMixed-signal Integrated CircuitReal Time MeasurementInstrumentationRadiation ImagingRadiation OncologyRadiologyHealth SciencesRadiation DetectionImplantable SensorRadiation MonitoringMonolithic 180MicroelectronicsDosimetryBiomedical SensorsRadiation DoseMedicineReal Time
The design and development of a monolithic system-on-chip silicon device for real time measurement of radiation dose is described. The device is fabricated in a standard 180 nm CMOS technology. The proposed application is in the medical field of In Vivo Dosimetry, where the dose of radiation, delivered to patients during radiotherapy sessions, is monitored in real time by an in-body implanted device of 1 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sup> approximate size and the related information is transmitted wirelessly to an external receiver operating in the Medical Implant Communication Service (MICS) band. Along with standard electrical characterization, evaluation of device performance is carried out under radiation beam in a clinical environment. The design and initial test results of the first phase of the project, covering the radiation sensors, will be discussed. A description of the next project steps will also be given.