Publication | Closed Access
Reduction of Magnetic Resonance Imaging-related Heating in Deep Brain Stimulation Leads Using a Lead Management Device
82
Citations
13
References
2005
Year
Small, concentric loops placed around the burr hole seem to reduce MRI-related heating for these implants. Although the mechanism is still not fully understood, a device such as that used in the present study could permit a wider range of clinical scanning sequences to be used at 1.5 and 3 T in patients with DBS implants, in addition to increasing the margin of safety for the patient.
| Year | Citations | |
|---|---|---|
Page 1