Accuracy of Measurements on Digital Videostroboscopic Images of the Vocal Folds

Renée Speyer, G.H. Wieneke, Willem Kersing, P.H.O. Dejonckere

Annals of Otology Rhinology & Laryngology · 2005 · 13 citations · 16 references

Concepts

Abstract

OBJECTIVES: The reliability of objective measurements on digital laryngeal images was investigated. METHODS: The magnitude of the error of measurement of surface areas by visually tracing the outline was determined for three different areas: the area of the lesion and, during vibration, the glottal area on maximal opening of the vocal folds and the glottal area on maximal closing of the vocal folds. RESULTS: The errors in these areas were 10% to 30%. The results suggest that the error is mainly due to the uncertainty of the real outline of areas with a vague boundary. Correction for differences in magnification between two images (posttherapy and pretherapy) is of importance in about 25% to 65% of cases, depending on the area measured. CONCLUSIONS: Only when the magnification ratio is small (less than about +/- 10% from 1.0) may a correction not be necessary.