Review of Scientific Instruments · 2012 · 16 citations · 7 references
MedicineRadiologyMetrologyUltrasonographyUltrasoundDistance MeasurementRange EstimationScanning TechniquesUltrasonicsUltrasound ImagingUltrasound InstrumentationB-scan ImagingComparative EvaluationLeast Squares EstimationMean Value Analysis
The aim of the present study is to develop and compare two different automatic methods for accuracy evaluation in ultrasound phantom measurements on B-mode images: both of them give as a result the relative error e between measured distances, performed by 14 brand new ultrasound medical scanners, and nominal distances, among nylon wires embedded in a reference test object. The first method is based on a least squares estimation, while the second one applies the mean value of the same distance evaluated at different locations in ultrasound image (same distance method). Results for both of them are proposed and explained.
7
Objective measurements of image quality
Jacinta E. Browne, Amanda J. Watson, N.M. Gibson et al. · Ultrasound in Medicine & Biology · 2004 · 66 citations · Full text
Arts, Radiology, Computer Vision +13