Medical Physics · 2000 · 27 citations · 9 references
MedicineRadiologyBiomedical ImagingDiagnostic ImagingTreatment PlanningComputed TomographyComputed Tomography PhysicsImage ReconstructionThree-dimensional ReconstructionCatheter ReconstructionComputed Tomography ReconstructionComputed Tomography-guided TherapyComputer-aided ReconstructionComputer-guided ReconstructionCt Image ReconstructionReconstruction AlgorithmsTreatment Planning Software
The aim of this study is to develop an automatic reconstruction of brachytherapy catheters using CT (computed tomography) data. Previously no such automatic facility has existed in any treatment planning software. To achieve this facility we have developed tools for the automatic reconstruction (which we term autoreconstruction) of plastic and metallic catheters. These algorithms overcome a number of difficulties which arise when a large number of catheters are present. These include situations with intersecting catheters and with loop techniques. The time required for the catheter reconstruction process using our autoreconstruction method is significantly reduced. The accuracy of our autoreconstruction is at least as high as the classical manual slice-by-slice method.
9
Frank A. Vicini, David A. Jaffray, Eric M. Horwitz et al. · International Journal of Radiation Oncology*Biology*Physics · 1998 · 71 citations
Sonja L. Schoeppel, Benedick A. Fraass, M P Hopkins et al. · International Journal of Radiation Oncology*Biology*Physics · 1989 · 58 citations