A capacitance-based micropositioning system for x-ray rocking curve measurements

G. L. Miller, R.A. Boie, P. L. Cowan, J. A. Golovchenko, R. W. Kerr, D. A. Robinson

Review of Scientific Instruments · 1979 · 14 citations · 5 references

DOIFull text

Open access

Concepts

Abstract

Certain types of x-ray experiments require very precise angular positioning of crystals in the vicinity of the Bragg reflection condition. A system applicable to some measurement of this type is described which achieves a long-term angular stability of approximately 10(-7) radians, coupled with a linear angular readout. This is achieved through a novel capacitance sensing system which provides angle measurement, together with the use of an auxilliary servo loop based on the Bragg condition to ensure long-term overall angular stability.

References