Journal of Analytical Atomic Spectrometry · 1989 · 33 citations · 0 references
Natural SciencesChemistryAnalytical ChemistryInorganic ChemistrySpectroscopyAbsorption SpectroscopyAtomic Absorption SpectroscopyAtomic SpectroscopyInorganic Elemental AnalysisInstrumental AnalysisMass SpectrometryCharacteristic MassGraphite Furnace Atomic Absorption SpectrometryGraphite Furnace Atomic Absorption SpectroscopyHydride Generation Atomic Absorption SpectrometryHydride Generation Atomic SpectroscopyInstrument Detection LimitMethod Detection LimitPre-concentration
A method is described for the determination of inorganic Ge in aqueous samples using hydride trapping and atomisation in a graphite furnace. Addition of Pd to the furnace tube surface before hydride trapping leads to large improvements in the sensitivity of the method. Extension of the procedure developed for Ge to As, Bi, Hg, Sb, Se, Sn and Te is demonstrated and the improvements in sensitivity with Pd are indicated. Analytical figures of merit including characteristic mass, precision and detection limit are given. The role of Pd is discussed.