Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 1989 · 25 citations · 0 references
MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImage ReconstructionImagingMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingArtifact ReductionArtifacts In Magnetic Resonance ImagingFourier ReconstructionImaging Signal ProcessingMagnetic Resonance Imaging AcquisitionResearch With Human Subjects
In MR imaging, the extent of the acquired spatial frequencies of the object is necessarily finite. The resulting image shows artefacts caused by "truncation" of its Fourier components. These are known as Gibbs artefacts or ringing artefacts. These artefacts are particularly. visible when the time-saving reduced acquisition method is used, say, when scanning only the lowest 70% of the 256 data lines. Filtering the data results in loss of resolution. A method is described that estimates the high frequency data from the low-frequency data lines, with the likelihood of the image as criterion. It is a computationally very efficient method, since it requires practically only two extra Fourier transforms, in addition to the normal. reconstruction. The results of this method on MR images of human subjects are promising. Evaluations on a 70% acquisition image show about 20% decrease of the error energy after processing. "Error energy" is defined as the total power of the difference to a 256-data-lines reference image. The elimination of ringing artefacts then appears almost complete..