Medical Physics · 2007 · 185 citations · 32 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyCardiac ImagingCardiovascular ImagingComputed TomographyComputed Tomography PhysicsImage ReconstructionMedical ImagingCt Image ReconstructionGatingIterative Image ReconstructionIterative ReconstructionMurine ModelsRegularized ReconstructionSparse ReconstructionTotal Variation Minimization
Recent advances in murine cardiac studies with three-dimensional (3D) cone beam micro-CT used a retrospective gating technique. However, this sampling technique results in a limited number of projections with an irregular angular distribution due to the temporal resolution requirements and radiation dose restrictions. Both angular irregularity and undersampling complicate the reconstruction process, since they cause significant streaking artifacts. This work provides an iterative reconstruction solution to address this particular challenge. A sparseness prior regularized weighted l2 norm optimization is proposed to mitigate streaking artifacts based on the fact that most medical images are compressible. Total variation is implemented in this work as the regularizer for its simplicity. Comparison studies are conducted on a 3D cardiac mouse phantom generated with experimental data. After optimization, the method is applied to in vivo cardiac micro-CT data.
32
David L. Donoho · IEEE Transactions on Information Theory · 2006 · 23.1K citations
L.A. Feldkamp, L. C. Davis, J. W. Kress · Journal of the Optical Society of America A · 1984 · 6.2K citations