Analytical Chemistry · 2002 · 57 citations · 17 references
Natural SciencesChemistryAnalytical ChemistryElemental AnalysisSpectroscopyParticle SizeParticulate MatterTrace Element AnalysisX-ray Fluorescence AnalysisX-ray SpectroscopyAirborne Particulate AnalysisAsh Composition AnalysisMicroscale Trace Element AnalysisParticulate Matter CharacterizationSimulation Model VerificationSynchrotron X-ray Spectroscopy
A new quantification procedure was developed for the evaluation of X-ray microfluorescence (XRF) data sets obtained from individual particles, based on iterative Monte Carlo (MC) simulation. Combined with the high sensitivity of synchrotron radiation-induced XRF spectroscopy, the method was used to obtain quantitative information down to trace-level concentrations from micrometer-sized particulate matter. The detailed XRF simulation model was validated by comparison of calculated and experimental XRF spectra obtained for glass microsphere standards, resulting in uncertainties in the range of 3-10% for the calculated elemental sensitivities. The simulation model was applied for the quantitative analysis of X-ray tube and synchrotron radiation-induced scanning micro-XRF spectra of individual coal and wood fly ash particles originating from different Hungarian power plants. By measuring the same particles by both methods the major, minor, and trace element compositions of the particles were determined. The uncertainty of the MC based quantitative analysis scheme is estimated to be in the range of 5-30%.
17
Analysis of X‐ray spectra by iterative least squares (AXIL): New developments
Bart Vekemans, Koen Janssens, László Vincze et al. · X-Ray Spectrometry · 1994 · 460 citations