Archives of Surgery · 1988 · 70 citations · 22 references
MedicineRadiologySurgeryDiagnostic ImagingDiagnostic RadiologySurgical PathologyBiopsy TechniquesBreast SurgeryForeign Body LocalizationForeign Body RemovalForeign Body RetrievalBiopsy LocalizationBiopsy Needle LocalizationBiopsy Needle PlacementIntraoperative Foreign Body LocalizationNonpalpable Breast Lesion LocalizationPercutaneous Foreign Body RetrievalWire Localization
Needle-wire localization of foreign bodies and nonpalpable breast lesions is commonly used to allow for more accurate excision or biopsy. We present three examples of complications of the localization procedure: (1) wire migration into the chest wall with retained fragment, (2) transection of a wire during biopsy with retained hook fragment, and (3) wire migration within the thigh soft tissues with breakage at the hooked end. Recommendations to minimize the incidence of these complications and their sequelae include (1) bending the hookwire 90 degree at the skin surface following localization, (2) transferring the patient between the radiology suite and the operating room via a stretcher, with minimal movement of the body part localized, and (3) accounting for the entire length of wire by the surgeon, pathologist, and radiologist following the procedure to exclude retained fragments.
22
Occult breast abnormalities: percutaneous preoperative needle localization.
Jack E. Meyer, D B Kopans, Paul C. Stomper et al. · Radiology · 1984 · 110 citations
Prebiopsy localization of nonpalpable breast lesions
JJ Gisvold, J P Martin · American Journal of Roentgenology · 1984 · 79 citations
Medicine, Surgery, Breast Oncology +15