Biomedical Chromatography · 2009 · 24 citations · 15 references
PharmacyPharmacokineticsPharmacologyAnalytical ToxicologyChromatographic AnalysisMass Spectrometric AnalysisBioanalytical Method ValidationChromatographyGradient ElutionLiquid ChromatographyLinear Dynamic RangeLinear RangeLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry AnalysisLiquid Chromatography-mass Spectrometry Method DevelopmentPrecisionTwo-dimensional Liquid ChromatographyTwo-dimensional Liquid Chromatography-mass Spectrometry
A highly sensitive and selective on-line two-dimensional reversed-phase liquid chromatography/electrospray ionization-tandem mass spectrometry (2D-LC-ESI/MS/MS) method was developed and validated to determine rifaximin in rat serum by direct injection. The 2D-LC-ESI/MS/MS system consisted of a restricted access media column for trapping proteins as the first dimension and a Waters C(18 )column as second dimension using 0.1% aqueous acetic acid:acetonitrile as mobile phase in a gradient elution mode. Rifampacin was used as an internal standard. The linear dynamic range was 0.5-10 ng/mL (r(2) > 0.998). Acceptable precision and accuracy were obtained over the calibration range. The assay was successfully used in analysis of rat serum to support pharmacokinetic studies.
15
Junmin Peng, Joshua E. Elias, Carson C. Thoreen et al. · Journal of Proteome Research · 2002 · 1.7K citations
Pharmacokinetic study of rifaximin after oral administration in healthy volunteers.
Jean-Jacques Descombe, D Dubourg, Matthieu Picard et al. · PubMed · 1994 · 124 citations