2012 · 14 citations · 9 references
EngineeringBiomedical EngineeringElectrical EngineeringBiomemsLow-power DesignWireless CommunicationsImplant DesignImplant TechnologyImplantable Biomedical DevicesImplantable MicrodevicesImplantable MicrosystemsWireless Power TransferBiomedical ImplantsBiomedical Wireless Power TransferElectromagnetic Field TheoryImplantable Wireless Power TransferMisalignment EffectsResonant Energy TransferResonant Power TransferWireless Power Transfer For Implants
We present a power transfer sensitivity analysis of a four-coil resonance-based wireless power delivery system that is used to provide energy to a small-size biotelemetry system implanted into non-human freely moving subjects. Both axial and angular misalignments have been evaluated, with calculations for the worst-case misalignment of 15mm axial distance and 50° angular separation for a minimum constant power supply of 0.5mW in the load resistor. Analytical results have been validated with electromagnetic field simulations. Discrete level design of a wirelessly powered implant incorporating this analysis is also presented.