Natural SciencesChemistryMaterials ScienceAnalytical ChemistryElemental AnalysisRadiation DetectionAlloysChemical Composition AnalysisElemental Composition AnalysisMetalsX-ray Photoelectron SpectroscopyAnalytical Detection LimitsAnalytical Method DevelopmentMetallurgical Elemental AnalysisQuantitative Analysis Of ElementsSample Analysis
Abstract An absolute quantitative analytical method without external standards has been developed for thick sample analysis by XRF and PIXE and has been applied to bronze and brass alloys. Comparisons between the results obtained and the chemical results were made and the detection limits for the determined elements were evaluated. In an Appendix a rapid calculation of the efficiency of the Ge(Hp) detector is presented.