Genosensor for SARS Virus Detection Based on Gold Nanostructured Screen‐Printed Carbon Electrodes

Graciela Martínez‐Paredes, Marı́a Begoña González-Garcı́a, Agustín Costa‐García

Electroanalysis · 2009 · 75 citations · 24 references

DOIFull text

Open access

Concepts

Abstract

Abstract A hybridization‐based genosensor is designed on a disposable gold nanostructured screen‐printed carbon electrode. Gold nanoparticles which are formed in situ by applying a constant current intensity during a fixed time, acts as an immobilization and transduction surface. The parameters affecting the methodology are studied, with special attention being placed on sensibility. Immobilization takes place through thiol‐gold interaction in a relatively fast way, and the genosensor response is found to be linearly to the biotinylated target concentration between 2.5 and 50 pmol/L; a detection limit of 2.5 pmol/L is calculated, and a sensitivity of 1.76 μA/pmol L −1 is achieved. These results improve significantly those described in literature, thus demonstrating the advantages offered by the use of gold‐nanostructured surfaces.

References

24