Publication | Closed Access
International comparisons of He-Ne lasers stabilized with<sup>127</sup>I<sub>2</sub>at = 633 nm (July 1993 to September 1995). Part II: Second comparison of Northern European lasers at = 633 nm
20
Citations
10
References
1997
Year
EngineeringLaser ScienceMeasurementBipm4 LaserLaser ApplicationsLaser PhysicsAccuracy And PrecisionLaser SimulationHigh-power LasersCalibrationInstrumentationInternational ComparisonsOptical PumpingPhotonicsNorthern European LasersPhysicsLaser Beam PropagationLaser DesignNatural SciencesSpectroscopyApplied PhysicsLaser SafetyGas LasersTunable LasersHigh-energy LasersAllan Standard DeviationsSeptember 1995Standard Uncertainty
This second Northern European comparison between seven lasers exhibits closer results than the first comparison. A standard uncertainty of 5 kHz was obtained against 15 kHz in 1990, while the frequency average of the group of lasers was about the same with 3,9 kHz, using the BIPM4 laser as reference. A large frequency difference on a laser observed during the first comparison was again observed; partial explanations of the origin of this difference were found. This difference has been excluded from the calculation of the average. Typical frequency stabilities using Allan standard deviations of about 1,6 × 10-11 and 1,7 × 10-12 were observed on 1s and 100 s sampling times, respectively.
| Year | Citations | |
|---|---|---|
Page 1
Page 1