GRAVITY: the calibration unit

Nicolas Blind, F. Eisenhauer, M. Haug, S. Gillessen, M. Lippa, L. Burtscher, O. Hans, Stefan Huber, A. Janssen, S. Kellner,

Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2014 · 13 citations · 4 references

DOIFull text

Open access

Concepts

Abstract

We present in this paper the design and characterisation of a new sub-system of the VLTI 2<sup><i>nd</i></sup> generation instrument GRAVITY: the Calibration Unit. The Calibration Unit provides all functions to test and calibrate the beam combiner instrument: it creates two artificial stars on four beams, and dispose of four delay lines with an internal metrology. It also includes artificial stars for the tip-tilt and pupil guiding systems, as well as four metrology pick-up diodes, for tests and calibration of the corresponding sub-systems. The calibration unit also hosts the reference targets to align GRAVITY to the VLTI, and the safety shutters to avoid the metrology light to propagate in the VLTI-lab. We present the results of the characterisation and validtion of these differrent sub-units.

References