Calibration of DC Voltage Dividers up to 100 kV

S. D'Emilio, Ferdinando Gabanna, G. La Paglia, Mauro Negro, Giancarlo Rua

IEEE Transactions on Instrumentation and Measurement · 1985 · 15 citations · 4 references

Concepts

Abstract

A system for calibrating dc voltage dividers up to 100 kV with an uncertainty (1α) of the order of 5 X 10-6 is described. The main component is a 200 MΩ guarded resistive device rated for the maximum voltage, forced-air cooled, and made up of 100 individually screened 2 MΩ resistors connected in series. An arrangement for series-to-parallel connection of the 100 resistors is provided in order to evaluate the total series resistance and the self-heating effect at full current and low voltage.

References

4