Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil
10
Citations
21
References
2015
Year
The agreement between both instruments was poor and the PIMA underestimated CD4 cell counts, which was more pronounced at CD4 counts ≥500 cell/μl. The PIMA's performance with fingerprick blood was less reliable than its performance with venous blood. In Brazil, where antiretroviral treatment is initiated regardless of CD4 counts, the PIMA's systematic bias towards CD4 underestimation may limit its role for monitoring HIV-patients.