Medical Physics · 2006 · 54 citations · 15 references
EngineeringBiomedical EngineeringBiophysicsBiomedical ImagingBiomedical OpticsDiagnostic ImagingBioluminescenceBioluminescence ImagingCellular ImagingDiffusion ApproximationLight-tissue InteractionMedical ImagingBioluminescence TomographyBorn ApproximationDiffusion Approximation Of Light TransportGreen's Function FormulationGreen's Function Representation FormulaNumerical SimulationSmall Animal ImagingTomographic Bioluminescence Imaging
In this paper, we present a Born-type approximation method for bioluminescence tomography (BLT), which is to reconstruct an internal bioluminescent source from the measured bioluminescent signal on the external surface of a small animal. Based on the diffusion approximation for the photon propagation in biological tissue, this BLT method utilizes the Green function to establish a linear relationship between the measured bioluminescent signal and the internal bioluminescent source distribution. The Green function can be modified to describe a heterogeneous medium with an arbitrary boundary using the Born approximation. The BLT reconstruction is formulated in a linear least-squares optimization framework with simple bounds constraint. The performance of this method is evaluated in numerical simulation and phantom experiments.
15
Optical tomography in medical imaging
Simon Arridge · Inverse Problems · 1999 · 2.5K citations
Alexander Sesonske · Nuclear Technology · 1976 · 948 citations
In vivo imaging of light-emitting probes
Brad Rice, M. D. Cable, Michael Nelson · Journal of Biomedical Optics · 2001 · 498 citations