Analytical Chemistry · 2002 · 134 citations · 16 references
Natural SciencesChemistryMaterials ScienceElectrochemical EngineeringElectrochemistryMicrofabricationElectrochemical CellsElectrode MaterialsElectrode PotentialsIon-selective ElectrodesWorking ElectrodesDissolved OxygenElectrode Response TimeMicroelectrode FabricationMicroelectrode MaterialsPh ElectrodesPh MonitoringRedox MediatorsResponse Time
A new method for fabrication of anodically electrodeposited iridium oxide film pH microelectrodes has been developed in this study. Novel for its tip size (3-10-microm tip diameter), the microelectrode is fabricated in a tapered glass micropipet filled with a low melting point alloy. The tapered end is recessed and platinized. Thereafter, iridium oxide is electrodeposited over the platinized end in the recessed part. The microelectrode has a very short response time (t80 < 5 s) in the pH range of 0-12 with an accuracy of 0.05 pH unit. The pH microelectrode is not affected by most ions and complexing agents of relevance in environmental and biological studies; it can be used in fluids over wide ranges of stirring speeds (0-55 rpm) and temperatures (approximately 5-40 degrees C). Redox agents such as dissolved oxygen and hydrogen peroxide have no effect on the pH response while quinhydrone, ferro- and ferricyanide, and sodium sulfide have marked effects. However, the microelectrode can still be used in any sample when calibration is done in standards having similar redox characteristics.
16
Xenia Beebe, Timothy L. Rose · IEEE Transactions on Biomedical Engineering · 1988 · 220 citations
Natural Sciences, Biomedical Engineering, Electrochemical Engineering +10
Electrochromism in anodic iridium oxide films
S. Gottesfeld, J. D. E. McIntyre, G. Beni et al. · Applied Physics Letters · 1978 · 175 citations