Natural SciencesBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Pulse SequencesMagnetic Resonance Signal FormationMedical ImagingMagnetic Resonance Imaging AcquisitionPulse Sequence DesignRadiofrequency Pulses
The method uses an actual flip‑angle imaging (AFI) pulse sequence with two identical RF pulses separated by two different delays (TR1 < TR2), acquiring a gradient‑echo signal after each pulse and operating in a pulsed steady‑state. A fast, image‑based method for measuring the transmitted RF field was developed, showing that the signal‑intensity ratio from two short TRs depends on flip angle, enabling accurate 3‑D B1 mapping that is insensitive to T1 and eliminates the need for long relaxation delays, thus providing time‑efficient whole‑body B1 mapping and correction of T1 maps under nonuniform RF excitation. Published in Magn Reson Med 57:192–200 (2007) and © 2006 Wiley‑Liss, Inc.
Abstract A new method has been developed for fast image‐based measurements of the transmitted radiofrequency (RF) field. The method employs an actual flip‐angle imaging (AFI) pulse sequence that consists of two identical RF pulses followed by two delays of different duration (TR 1 < TR 2 ). After each pulse, a gradient‐echo (GRE) signal is acquired. It has been shown theoretically and experimentally that if delays TR 1 and TR 2 are sufficiently short and the transverse magnetization is completely spoiled, the ratio r = S 2 / S 1 of signal intensities S 1 and S 2 , acquired at the beginning of the time intervals TR 1 and TR 2 , depends on the flip angle (FA) of applied pulses as r = (1 + n * cos(FA))/( n + cos(FA)), where n = TR 2 /TR 1 . The method allows fast 3D implementation and provides accurate B 1 measurements that are highly insensitive to T 1 . The unique feature of the AFI method is that it uses a pulsed steady‐state signal acquisition. This overcomes the limitation of previous methods that required long relaxation delays between sequence repetitions. The method has been shown to be useful for time‐efficient whole‐body B 1 mapping and correction of T 1 maps obtained using a variable FA technique in the presence of nonuniform RF excitation. Magn Reson Med 57:192–200, 2007. © 2006 Wiley‐Liss, Inc.
28
Imaging of the active B1 field in vivo
Rudolf Stollberger, Paul Wach · Magnetic Resonance in Medicine · 1996 · 355 citations