Asymmetric response of superconducting niobium-tunnel-junction x-ray detectors

M. Ohkubo, Jens Martin, K. Drachsler, R. P. Huebener, I. Sakamoto, Neisei Hayashi

Physical review. B, Condensed matter · 1996 · 16 citations · 37 references

Concepts

Abstract

The asymmetric response of the counter and base electrodes of x-ray detectors employing polycrystalline niobium tunnel junctions has been studied by low-temperature scanning electron microscopy (LTSEM). LTSEM has revealed that the base electrode produces a signal more than two times larger than the counter electrode, and effective quasiparticle lifetimes are 100 ns in the counter and 280 ns in the base. Based on the I-V characteristics and the measured effective quasiparticle lifetimes, we propose a model structure for a spatial variation of the gap parameter \ensuremath{\Delta}, which involves proximity quasiparticle trapping layers. The small counter signal is caused by the shorter quasiparticle lifetime and the trapping effect. The LTSEM results are consistent with x-ray spectra for a radioactive $^{55}\mathrm{Fe}$ source. \textcopyright{} 1996 The American Physical Society.

References

37