Optics Express · 2011 · 41 citations · 9 references
EngineeringBioengineeringBiomedical EngineeringBiomedical OpticsPhotonicsSpectroscopyClinical BiosensingOptical BiosensingOptical DiagnosticsOptical SensingOptical SpectroscopyRaman SpectroscopyBiosensingHollow-core FibersLaser-based BiosensingOptical DesignOptical DetectionRaman Spectroscopy SensingReal-time MonitoringResponse Time
The feasibility of using hollow core photonic crystal fiber (HC-PCF) in conjunction with Raman spectroscopy has been explored for real time monitoring of heparin concentration in serum. Heparin is an important blood anti-coagulant whose precise monitoring and controlling in patients undergoing cardiac surgery and dialysis is of utmost importance. Our method of heparin monitoring offers a novel alternative to existing clinical procedures in terms of accuracy, response time and sample volume. The optical design configuration simply involves a 785-nm laser diode whose light is coupled into HC-PCF filled with heparin-serum mixtures. By non-selectively filling HC-PCF, a strong modal field overlap is obtained. Consequently, an enhanced Raman signal (>90 times) is obtained from various heparin-serum mixtures filled HC-PCFs compared to its bulk counterpart (cuvette). The present scheme has the potential to serve as a 'generic biosensing tool' for diagnosing a wide range of biological samples.