Measurement Science and Technology · 2007 · 25 citations · 19 references
EngineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsOpticsFourier OpticsWavefront-guided CorrectionEntry AngleFringing FieldsRepositioningTermination Conditions
Strong fringing fields at the entry and exit of a real hemispherical deflector analyser (HDA) significantly degrade the 180° first-order focusing conditions, one of the central advantages of the ideal-field HDA. Over the past 50 years, traditional approaches to cure this problem have primarily sought to suppress these fields by improving field termination conditions typically requiring the unwieldy use of additional electrodes. Recently, Zouros et al (2006 Meas. Sci. Technol. 17 N81–N86) have shown in simulation that a simple repositioning of the HDA entry when appropriately biased results in the effective utilization of the intrinsic lensing properties of these fields to restore and even improve first-order focusing. Here, we investigate in simulation the efficacy of the new controlled lensing approach and compare it to the traditional Herzog and Jost field corrector approach. HDA focusing properties and energy resolution are reported as a function of entry angle, source extent and hemispherical interelectrode separation. For all cases considered, HDAs using controlled lensing always came out ahead demonstrating superior focusing along the 180° deflection plane and improved energy resolution.
19
Angle-resolved x-ray photoelectron spectroscopy
C. S. Fadley · Progress in Surface Science · 1984 · 640 citations
Design of electron spectrometers
Denis Roy, Daniel Tremblay · Reports on Progress in Physics · 1990 · 111 citations
Natural Sciences, Physics, Optics +12