Natural SciencesAstronomyInstrumentationRadiation DetectionSpectroscopyDetector TechnologyRadiation SpectroscopyX-ray DetectionX-ray DetectorsX-ray SpectroscopyCryogenic InstrumentationCryostat DesignDetector Energy ResolutionHigh-resolution X-ray SpectroscopyMicrocalorimeter DetectorsRadiation Energy ResolutionSoft X-ray SpectroscopySpectrometer Resolution
We present the development status of the Soft X-ray Spectrometer (SXS) onboard the ASTRO-H mission. The SXS provides the capability of high energy-resolution X-ray spectroscopy of a FWHM energy resolution of < 7eV in the energy range of 0.3 – 10 keV. It utilizes an X-ray micorcalorimeter array operated at 50 mK. The SXS microcalorimeter subsystem is being developed in an EM-FM approach. The EM SXS cryostat was developed and fully tested and, although the design was generally confirmed, several anomalies and problems were found. Among them is the interference of the detector with the micro-vibrations from the mechanical coolers, which is the most difficult one to solve. We have pursued three different countermeasures and two of them seem to be effective. So far we have obtained energy resolutions satisfying the requirement with the FM cryostat.
3
The ASTRO-H X-ray astronomy satellite
Tadayuki Takahashi, Kazuhisa Mitsuda, Richard L. Kelley et al. · Maryland Shared Open Access Repository (USMAI Consortium) · 2014 · 75 citations · Full text
Natural Sciences, Astronomy, Physics +15