Abstract

The purpose of this study was to assess the suitability of different radiation dosimetry methods and record radiation exposures during paediatric catheterization. Three methods of dosimetry were employed: thermoluminescent dosimetry (TLD), dose-area product and calculation of entrance surface dose from calibrated exposure factors. Examinations included bi-plane fluoroscopy, and cineangiography for diagnosis and treatment of congenital heart disease. The most suitable method of radiation dosimetry for cardiac catheterization is the use of calculated entrance dose or a dose-area product meter. Children were exposed to high levels of radiation during cardiac catheterization but there was a wide variation in radiation dosage. Careful consideration should be given to the suitability of radiation dosimetry for cardiac catheterization.