Publication | Closed Access
Radiation exposure to children during cardiac catheterization.
66
Citations
9
References
1997
Year
Radiation EffectRadiation ExposurePaediatric RadiologyRadiation ProtectionRadiation MedicinePediatric DosimetryCardiologyCongenital Heart DiseaseRadiologyHealth SciencesIonizing RadiationRadiation DosimetryRadiation SafetyRadiation EffectsDosimetryCardiac CatheterizationPediatricsRadiation DoseMedicine
The purpose of this study was to assess the suitability of different radiation dosimetry methods and record radiation exposures during paediatric catheterization. Three methods of dosimetry were employed: thermoluminescent dosimetry (TLD), dose-area product and calculation of entrance surface dose from calibrated exposure factors. Examinations included bi-plane fluoroscopy, and cineangiography for diagnosis and treatment of congenital heart disease. The most suitable method of radiation dosimetry for cardiac catheterization is the use of calculated entrance dose or a dose-area product meter. Children were exposed to high levels of radiation during cardiac catheterization but there was a wide variation in radiation dosage. Careful consideration should be given to the suitability of radiation dosimetry for cardiac catheterization.
| Year | Citations | |
|---|---|---|
Page 1
Page 1