Journal of Physics Conference Series · 2010 · 51 citations · 8 references
Natural SciencesMaterials SciencePhysicsAnalytical ChemistryPlasma PhysicsRadiation MeasurementElectron MicroscopyTransmission Electron MicroscopyCarbon RemovalDark-field ImagingElectron Beam Irradiation EffectsScanning Electron MicroscopyScanning Transmission Electron MicroscopySurface ContaminationSurface Contamination MeasurementSurface Contamination Removal
We have measured the build-up of carbon surface contamination as a function of time and irradiated area size for various specimens in a JEOL 2010F (scanning) transmission electron microscope, employing both t/λ mapping with our Gatan imaging filter and recording changes in annular dark-field image intensity. It is shown that the total number of carbon atoms deposited per time for a given beam intensity is roughly constant at room temperature for as-received specimens while it is significantly lower for plasma cleaned specimens. This explains why contamination is generally only an issue at the highest magnifications where the contamination regions become smaller and the carbon layers correspondingly thicker. A Fischione plasma cleaner was then used to remove these carbon layers, and the rate of carbon removal has been determined for contamination spots produced in stationary spot mode as well as for extended regions scanned for a minute so that optimal cleaning times can be chosen.
8
Electron energy-loss spectroscopy
R.F. Egerton · Physics World · 1997 · 437 citations
Natural Sciences, Chemistry, Physics +14
Plasma Cleaning and Its Applications for Electron Microscopy
Thomas Isabell, Paul Fischione, Catherine A. O’Keefe et al. · Microscopy and Microanalysis · 1999 · 128 citations