Journal of Analytical Atomic Spectrometry · 1989 · 40 citations · 0 references
Natural SciencesChemistryAnalytical ChemistryElemental AnalysisSpectroscopyAtomic Absorption SpectroscopyHeavy Metal AnalysisInorganic Elemental AnalysisInstrumental AnalysisTrace AnalysisTrace Element AnalysisAnalytical Detection Of Heavy MetalsAtomic Fluorescence SpectroscopyBlank MeasurementDetection Limit AssessmentFlame Atomic Absorption SpectrometryFlame Atomic Absorption SpectroscopyMethod Detection LimitRoom Temperature Stability
A volatile cadmium compound can be generated from aqueous solutions containing Cd2+ at trace levels, by reaction with sodium tetraethylborate(III) at room temperature. Preliminary experiments failed to identify this compound but it appears to be sufficiently stable to be used for analytical purposes. A detection limit for the compound of 0.2 ng ml–1(3σ of a blank determination, 1-ml sample volume) can be achieved using non-dispersive atomic fluorescence spectrometry with argon-hydrogen mini-flame atomisation whereas with conventional atomic absorption spectrometric measurements in a flame-heated silica tube the detection limit is 1 ng ml–1(3δ of the base-line noise, 1-ml sample volume).