Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2007 · 20 citations · 9 references
EngineeringBioengineeringBiotechnologyBiomemsMicrofabricationNucleic Acid BiochemistryMicrofluidic DevicesMicrofluidic SystemsMicrofluidic TransportMicrofluidicsPolymerase Chain ReactionContinuous FlowContinuous Flow ProcessingContinuous-flow MicrofluidicsExperimental Results InterpretationMicrofluidic AnalysisMicrofluidic Chip DesignMicrofluidic FlowMicrofluidic Platforms
PCR is the most commonly used method to identify DNA segments. Several methods have been used to miniaturize PCR and perform it in a microfluidic chip. A unique approach is the continuous-flow PCR, where the conventional thermocycling is replaced by pumping the sample through a channel which meanders over stationary temperature zones, allowing fast temperature changes in the sample due to the low thermal mass as well as a continuous production of PCR products. In this paper we present a system comprising a polymer microfluidic chip, a thermocycler unit and the protocols necessary to perform fast continuous-flow PCR including experimental results in comparison with a conventional PCR system.
9
Chemical Amplification: Continuous-Flow PCR on a Chip
Martin U. Kopp, Andrew J. de Mello, A. Manz · Science · 1998 · 1.2K citations
PDMS-based micro PCR chip with Parylene coating
Young Shik Shin, Keunchang Cho, Sun Hee Lim et al. · Journal of Micromechanics and Microengineering · 2003 · 328 citations