IEEE Transactions on Biomedical Circuits and Systems · 2014 · 53 citations · 22 references
EngineeringBioengineeringBiomedical EngineeringAnalog ElectronicsBiomedical Signal ProcessingBiosensorsBiomedical InstrumentationCardiovascular MonitoringElectrocardiographyMedical InstrumentationPhysiological Signal AcquisitionWearable Health TechnologiesAnalog Front-end CircuitsCardiac Monitoring SystemsFerroelectric Random-access MemoryHeart RateHeart Rate MonitoringPhysiological MonitoringSignal-to-noise RatioWearable Health Devices
To prevent lifestyle diseases, wearable bio-signal monitoring systems for daily life monitoring have attracted attention. Wearable systems have strict size and weight constraints, which impose significant limitations of the battery capacity and the signal-to-noise ratio of bio-signals. This report describes an electrocardiograph (ECG) processor for use with a wearable healthcare system. It comprises an analog front end, a 12-bit ADC, a robust Instantaneous Heart Rate (IHR) monitor, a 32-bit Cortex-M0 core, and 64 Kbyte Ferroelectric Random Access Memory (FeRAM). The IHR monitor uses a short-term autocorrelation (STAC) algorithm to improve the heart-rate detection accuracy despite its use in noisy conditions. The ECG processor chip consumes 13.7 μA for heart rate logging application.