Review of Scientific Instruments · 1983 · 17 citations · 17 references
PharmacyBiochemistryBiophysicsLipid BiochemistryThermal AnalysisThermodynamicsCalorimetryEnthalpy MeasurementHeat Capacity MeasurementLipid AnalysisPhase Transition AnalysisTemperature MeasurementCalorimetry In Biological SystemsDirect CalorimetryHeat CapacityHeat Capacity DeterminationHeat Capacity EstimationLiquid Heat Capacity MeasurementPhase Transition ThermodynamicsSpecific Heat Determination
An ac calorimeter is developed for the measurement of heat capacity in liquid including the aqueous suspension of biological materials in the temperature range between 270 and 400 K. Using this apparatus we perform the measurement of heat capacity at the main phase transition of lipid suspended in water. Additionally a new method to determine the absolute value of the heat capacity is proposed.
17