IEEE Transactions on Nuclear Science · 1979 · 45 citations · 4 references
Veterinary MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyComputed TomographyComputed Tomography Dose ManagementComputed Tomography PhysicsMedical ImagingMedical Imaging PhysicsRadiation Imaging SystemsDetector Spatial ResolutionMultiple Coulomb ScatteringMultiple ScatteringSpatial Resolution MeasurementX-ray Imaging PhysicsX-ray Imaging Systems
The use of protons or other heavy charged particles instead of x rays in computed tomography (CT) is explored. The results of an experimental implementation of proton CT are presented. High quality CT reconstructions are obtained at an average dose reduction factor compared with an EMI 5005 x-ray scanner of 10:1 for a 30-cm-diameter phantom and 3.5:1 for a 20-cm diameter. The spatial resolution is limited by multiple Coulomb scattering to about 3.7 mm FWHM. Further studies are planned in which proton and x-ray images of fresh human specimens will be compared. Design considerations indicate that a clinically useful proton CT scanner is eminently feasible.
4
The Fourier reconstruction of a head section
L. A. Shepp, B. F. Logan · IEEE Transactions on Nuclear Science · 1974 · 2.2K citations
Detectability in computed tomographic images
Kenneth M. Hanson · Medical Physics · 1979 · 193 citations
The Application of Protons to Computed Tomography
K. M. Hanson, J. N. Bradbury, T. M. Cannon et al. · IEEE Transactions on Nuclear Science · 1978 · 37 citations
Proton Radiography in the Diagnosis of Breast Carcinoma
V. W. Steward, Andreas Koehler · Radiology · 1974 · 23 citations