Evaluation of acetabular cup initial fixation by using resonance frequency principle

Petr Henyš, Lukáš Čapek, Jaroslav Fencl, E Procházka

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · 2015 · 26 citations · 19 references

Concepts

Abstract

The clinical practice shows that the loosening of acetabular cups is more frequent than stem loosening. With standard cups, the incidence of dislocation failure is highest in the first year after arthroplasty implantation. The aim of the study was to quantitatively evaluate the implant-bone stability of a cementless acetabular cup prosthesis by using a device based on resonance frequency analysis. The evaluation of this device was done by finite element analysis and in vitro experiments. It was shown that not all the resonance frequencies can be measured by our device. The resonance frequencies vary within the range of 500-3000 Hz. The proposed power spectrum measurement gives the information about the absolute stiffness of the press-fit implant.