EngineeringNuclear EngineeringPhysicsNuclear InstrumentationNuclear PhysicsRadiation DetectionAccelerator PhysicsCharged Particle DetectionDetector TechnologyPulse ProcessingRadiation DetectorsRadiation SpectroscopyAcceleratorsDetector DesignPulse Height AnalysisPulse Height SpectraSampler InstrumentsSignal FormationSignal Processing For DetectorsVoltage
The energy of an electrostatic accelerator is modulated over a range of 50 kev by applying an alternating potential of 50-kv amplitude to an insulated target at a frequency of 10 cps. A sample of the modulating voltage and the signals from a nuclear detector are fed to a so-called ``function sampler,'' which furnishes an output pulse whose height is proportional to the instantaneous value of the target voltage. Analysis of the pulse-height spectrum by means of a 100-channel analyzer then affords a presentation of the spectrum of nuclear events as a function of accelerator energy over a 50-kev range, each channel of the analyzer corresponding to an energy width of 500 ev. Advantages of this system are discussed for high resolution work with an electrostatic accelerator.