Applied Optics · 2005 · 112 citations · 12 references
Natural SciencesAstronomyAtmospheric RadiationRadiation MeasurementSpectroscopyRadiation InstrumentationSolar IrradianceSolar RadiationSpectral RadiationUltraviolet InstrumentationUltraviolet SpectroscopyIrradianceSolar Irradiance MeasurementSolar Radiation MeasurementSolar Ultraviolet IrradianceSolar Zenith AngleSpectral Solar IrradianceUltraviolet Radiation MeasurementUltraviolet Solar Irradiance
A transportable reference spectroradiometer for measuring spectral solar ultraviolet irradiance has been developed and validated. The expanded uncertainty of solar irradiance measurements with this reference spectroradiometer, based on the described methodology, is 8.8% to 4.6%, depending on the wavelength and the solar zenith angle. The accuracy of the spectroradiometer was validated by repeated site visits to two European UV monitoring sites as well as by regular comparisons with the reference spectroradiometer of the European Reference Centre for UV radiation measurements in Ispra, Italy. The spectral solar irradiance measurements of the Quality Assurance of Spectral Ultraviolet Measurements in Europe through the Development of a Transportable Unit (QASUME) spectroradiometer and these three spectroradiometers have agreed to better than 6% during the ten intercomparison campaigns held from 2002 to 2004. If the differences in irradiance scales of as much as 2% are taken into account, the agreement is of the order of 4% over the wavelength range of 300-400 nm.